| Size (H x W x
D): |
17.0 x 11.5 x
10.5 in. (43 x 29 x 26.7 cm) (without power source) |
| Weight: |
17.3 lbs (7.8
kg) (without power source) |
| Material: |
High-impact
molded polystyrene structural foam |
| Enclosure: |
NEMA 4X (iP65) |
| Program Memory: |
Non-volatile,
programmable flash. Updateable without opening enclosure. |
| Display: |
Backlit LCD,
2-line, 80-character (5.5 mm high x 3.2 mm wide) |
| Storage
Temperature: |
-40° to 140°F
(-40° to 60°C) |
| Operating
Temperature: |
0° to 140°F
(-18° to 60°C) |
| Power: |
12 to 14V DC, 24
mA average at 12.5V DC |
| Typical
battery life (printer set at 1 in/hr, 1 minute level reading interval) |
| 934 NiCad
Battery: |
8 to 11 days |
| 946 Lead-Acid
Battery: |
12 to 16 days |
| 948 Lead-Acid
Battery: |
21/2 to 3 months |
| Submerged
probe |
| Dimensions: |
9.5 x 0.875 in
(24.1 x 2.2 cm) |
| Cable: |
Standard range
probe - 25 ft x 0.3 in (7.6 m x 0.8 cm); Extended range probe - 50 ft x
0.3 in (15.2 m x 0.8 cm) |
| Level
Measurement Method: |
Submerged
pressure transducer mounted in the flow stream |
| Transducer Type: |
Differential
linear integrated circuit pressure transducer |
| Hazardous
Location Rating: |
UL Classified
for Class I, Division 1, Groups A, B, C, & D locations, when
installed with Isco Intrinsically Safe Barrier and Quick Disconnect Box
per control drawing 60-3403-131 |
| Operating
Temperature: |
32° to 160°F
(0° to 71°C) |
| Compensated
Temperature: |
32° to 100°F
(0° to 38°C) |
| Level
Measurement Range |
| Standard rangep
probe: |
0.1 to 10 ft
(0.03 to 3.05 m) |
| Extended range
probe: |
0.1 to 30 ft
(0.03 to 9.14 m) |
| Maximum
Allowable Level: |
Standard range
probe - 20 ft (6.1 m); Extended range probe - 40 ft (12.2 m) |
| Level
Measurement Accuracy (at 25°C (77°F) - Max error (±) for indicated
level range) |
| Standard
range probe |
| 0.1 to 5.0 ft
(0.03 to 1.52 m): |
±0.01 ft (±0.003
m) |
| 0.1 to 7.0 ft
(0.03 to 2.13 m): |
±0.03 ft (±0.009
m) |
| 0.1 to 10 ft
(0.03 to 3.05 m): |
±0.10 ft (±0.03
m) |
| Extended
range probe |
| 0.1 to 15 ft
(0.03 to 4.57 m): |
±0.03 ft (±0.009
m) |
| 0.1 to 21 ft
(0.03 to 6.40 m): |
±0.09 ft (±0.027
m) |
| 0.1 to 30 ft
(0.03 to 9.14 m): |
±0.30 ft (±0.09
m) |
| Temperature
Coefficient (Maximum error per degree of temperature change over
compensated range) |
| Standard
range probe |
| 0.1 to 4.0 ft
(0.03 to1.22m): |
±0.005ft/°F (±0.0027m/°C) |
| 4.0 to 10 ft
(1.22 to 3.05m): |
±0.007ft/°F (±0.0038m/°C) |
| Extended
range probe |
| 0.1 to 30 ft
(0.03 to 9.14 m): |
±0.008ft/°F (±0.0044
m/°C) |
| Materials |
| Submerged probe: |
Type 316
stainless steel, chlorinated polyvinyl chloride (CPVC) |
| Cable: |
Polyvinyl
chloride (PVC) |